Z-2534-2024: Philips Azurion Interventional Fluoroscopic X-ray System with Software Version Number: R1.0, R1.1, R1.2-

Recall record Z-2534-2024 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2534-2024
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Aug 8, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential Loss of Imaging (X-ray) Functionality and/or Longer Time to Perform Cold Restart

Source facts

Action
Philips issued URGENT Medical Device Correction Letter to US consignees on July 15th, 2024, via certified mail. Letter states reason for recall, health risk and action to take: Circulate this Urgent Medical Device Correction to all users of the system so that they are aware of the issue. " Keep this Urgent Medical Device Correction with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. " Should you experience a loss of imaging (X-ray) functionality or a prolonged system restart time, call your local Philips representative to report the event. " Complete and return the attached response form (page 4) to Philips promptly and no later than 30 days from receipt. This confirms that the users of the system have reviewed and understand this Urgent Medical Device Correction and required actions to be taken. Philips is working on a software release (R2.2.10) that will correct these two issues (ref. FCOs: FCO72200548, FCO72200580, FCO72200582, FCO72200583, FCO72200584, FCO72200592). For those systems that have an Interventional Workspot (IW) and/or EchoNavigator, Philips will upgrade the software version of the IW and/or EchoNavigator to maintain compatibility with the updated Philips Azurion system software (R2.2.10). Your local Philips representative will contact you to schedule a visit to install the software update once available. Philips expects this software to be released by Q1 2025. If you need additional information or support concerning these issues, contact your local Philips representative at 1-800- 722-9377 (Philips Customer Care Solutions Center).
Initiated
2024-07-15
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan Palestine, State of Peru, Philippines, Poland, Portugal, R¿union Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam.
Recall number
Z-2534-2024
Recall status
Open, Classified
CDRH record ID
209082
FDA classified
2024-08-08
Recall event number
94994

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗