Open original FDA record ↗Recall record
- FDA identifier
- Z-2535-2017
- Company in FDA record
- Cook Inc.
- FDA record creation date
- Jun 14, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products
Source facts
- Action
- On 4/25/2017, URGENT: MEDICAL DEVICE RECALL notifications were sent to the affected consignees via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.
- Initiated
- 2017-04-25
- Root cause
- Labeling design
- Terminated
- 2019-03-26
- Distribution
- The devices have been distributed nationwide within the United States and the following countries: Australia, Belgium, Canada, Finland, Germany, Guatemala, Hong Kong, India, Ireland, Japan, South Korea, Kuwait, Malaysia, Netherlands, New Zealand, Puerto Rico, Saudi Arabia, Taiwan, and the United Kingdom.
- Recall number
- Z-2535-2017
- Recall status
- Terminated
- CDRH record ID
- 155459
- FDA record created
- 2017-06-14
- Recall event number
- 77217