The Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for delay in surgery if additional cement needs to be prepared for the injection procedure.
Source facts
Action
On 6/13/2016, URGENT MEDICAL DEVICE RECALL NOTIFICATION notifications were sent to the affected customers via courier and sales representatives via email. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments:
Jennifer Olson
269-389-2644
jennifer.olson@stryker.com
Initiated
2016-06-14
Root cause
Process control
Terminated
2017-03-27
Distribution
Worldwide distribution. US nationwide, France, Netherlands, Spain, Italy, Canada, Israel, Romania, Switzerland, and United Kingdom.
Recall number
Z-2536-2016
Recall status
Terminated
CDRH record ID
147944
FDA classified
2016-08-15
Recall event number
74582