Skip to content

Public FDA record

Z-2536-2020: The CapsoCam Plus (SV-3); Device Common Name: System, Imaging, Gastrointestinal, Wireless, Capsule; Catal

Recall record Z-2536-2020 for Capso Vision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2536-2020
Company in FDA record
Capso Vision, Inc.
FDA record creation date
Jul 8, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ingestible video capsule system has capsule with incorrect capture mode which will not record data when the capsule is ingested by patients. If used, there is a risk of delay of diagnosis and the patient may need to repeat the exam.

Source facts

Action
Beginning 02/26/2020, the firm notified affected customers via email of "URGENT: Medical Device Voluntary Recall Letter". The letter indicated, as soon as the customer receives the notice, please provide acknowledgment of receipt. The letter requested actions to be taken by the customer/user: We request that you ship back the affected capsules using the provided RMA form and return shipping label. For any concerns or question email or contact the firm at (408) 866-6358.
Initiated
2020-02-26
Root cause
Employee error
Terminated
2021-01-28
Distribution
US Nationwide distribution including in the states of AZ, CA, DE, MD, NJ, NY, SC, and VA.
Recall number
Z-2536-2020
Recall status
Terminated
CDRH record ID
181205
FDA record created
2020-07-08
Recall event number
85540

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗