Open original FDA record ↗Recall record
- FDA identifier
- Z-2537-2014
- Company in FDA record
- Smiths Medical ASD, Inc.
- FDA classification date
- Aug 28, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The tube fastener on the ET Tube Holder may compress the tracheal tube when used which may cause difficulty when locking the ET Tube Holder in place or difficulty when passing a suction catheter.
Source facts
- Action
- Smiths Medical sent an Urgent Medical Device Recall Notice dated July 11, 2014, to all affected customers. The letter informed customers of the recall and instructed them to inspect their inventory and segregate any unused affected products and to complete and return the Recall Confirmation Form, by Fax 1-800-621-2590 or email to endotrachealtubeholders@smiths-medical.com. Customers with questions were instructed to call Smiths Medical's Customer Service Department at 1-800-258-5361. For questions regarading this recall call 866-216-8806.
- Initiated
- 2014-07-07
- Root cause
- Process change control
- Terminated
- 2021-05-26
- Distribution
- Worldwide Distribution - USA including AK, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to CA, CY, GB, IE, IT, JE, KW, MT, PL, RU, SA, ZA.
- Recall number
- Z-2537-2014
- Recall status
- Terminated
- CDRH record ID
- 128844
- FDA classified
- 2014-08-28
- Recall event number
- 68739