Open original FDA record ↗Recall record
- FDA identifier
- Z-2537-2018
- Company in FDA record
- Forte Automation Systems Inc
- FDA record creation date
- Jul 25, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Communications between the Patient Positioning System and the accuracy filter can periodically fail with no clear indication to the operator.
Source facts
- Action
- On September 5, 2017 Forte has distributed an Urgent Medical Device Correction letter with the following information, Subject: Robotic Treatment Table Positioning Error Commercial Name of Affected Product: Patient Positioning System Affected Version(s) / Lot(s): KRC2 Controllers with software version 2.3.1 2.3.10 Reference / FSCA Identifier: 3003900998-2017-00001 Type of Action: Notification and Correction This letter provides users with a Description of the Problem, Details, Recommended User Action and Forte Automation Systems, Inc. Actions. This document also Pending resolution with a permanent solution, Forte Automation Systems, Inc. recommends reimaging the patient prior to treatment, especially after any table movement. This includes movement initiated by the image position verification system. It also advises users that a solution for the issue is in development.
- Initiated
- 2017-09-05
- Root cause
- Software design
- Terminated
- 2020-04-13
- Distribution
- US in the state of California
- Recall number
- Z-2537-2018
- Recall status
- Terminated
- CDRH record ID
- 165646
- FDA record created
- 2018-07-25
- Recall event number
- 80510