Z-2537-2020: VistaSolution, physiological signal transmitter. A software graphical user interface for use by healthca

Recall record Z-2537-2020 for Vitalconnect Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2537-2020
Company in FDA record
Vitalconnect Inc.
FDA classification date
Jul 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but no notification was found. A software defect was encountered causing the lack of notification. The defect could result in exposing a patient to potential hypoxia for a short window of time. The firm is correcting the software to bring the performance back into compliance with specifications.

Source facts

Action
On May 21, 2020, the firm sent a "Urgent: Medical Device Recall" notification to all affected customers via E-mail and mail. The letter stated, that the third party Sp02 continuous device notifications may not indicate an alert as intended. Therefore, the health care provider may not be notified to check the patient. Potential Hazards: A patient may have hypoxia and the designated health care provider may not be notified. The firm will be turning off the ability to set custom notifications containing continuous SP02 until the issue is resolved. The firm tells customers to monitor the patients for SPO2, as if the patients are not connected to the recalled device. The customers are informed to acknowledge the receipt of this notification by E-mail or via mail to the firm, email to: cmerrell@vitalconnect.com and if any questions, then contact: Tony Wilder at phone: (408) 963-4600 or support@vitalconnect.com
Initiated
2020-05-21
Root cause
Software Design Change
Terminated
2021-05-13
Distribution
U.S. Nationwide distribution including in the states of NJ, NY, VA, OUS: None
Recall number
Z-2537-2020
Recall status
Terminated
CDRH record ID
181805
FDA classified
2020-07-10
Recall event number
85802

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗