Skip to content

Public FDA record

Z-2537-2024: Hookup Model/Catalog Number: 2-8-540HAN Product Description: Hookups provide the interface between an Ad

Recall record Z-2537-2024 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2537-2024
Company in FDA record
Steris Corporation
FDA classification date
Aug 8, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection

Source facts

Action
Steris issued Urgent Medical Device Correction letter on 7/10 via FedEx mail. Letter states reason for recall, health risk and action to take: User Action Please ensure the following steps are completed: 1.Locate and destroy your IFU(s) relative to the hookup(s) affected by this recall. 2.Complete the Medical Device Recall Response Form included with this letter. 3.Return the completed Response Form via email to: Regulatory_Compliance@STERIS.com or via fax to 440-392-8963. As you are familiar, users have the ability to update hookup and parameter set settings in the Advantage AERs to correspond with your inventory of endoscopes. For your convenience, we have included an excerpt from the Operator Manual (Advantage Plus Pass-Thru, pages 138-140, and Advantage Plus, page 72) of the instructions on how to make these updates (see Attachment A or B, as applicable). Users are reminded to report any adverse events via FDA s MedWatch Database should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact your local STERIS Representative or STERIS Technical Support at 800-548-4873 or 800-444-4729.
Initiated
2024-06-13
Root cause
Error in labeling
Distribution
Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Malaysia, Qatar, Ukraine, United Arab Emirates, Nicaragua, and Nigeria.
Recall number
Z-2537-2024
Recall status
Open, Classified
CDRH record ID
209036
FDA classified
2024-08-08
Recall event number
94987

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗