Skip to content

Public FDA record

Z-2539-2017: Cook Multi-Use Holmium Laser Fibers, Supplied sterilized by ethylene oxide gas in peel-open packages

Recall record Z-2539-2017 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2539-2017
Company in FDA record
Cook Inc.
FDA classification date
Jun 5, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Correction to the product Instructions for Use because the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.

Source facts

Action
Cook is initiating a voluntary recall of multiple products in the form of a correction to the product Instructions for Use. Cook Inc. has identified the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products. Potential adverse events that may occur if the products are not adequately reprocessed include urological infections and systemic infections from a urological origin as well as events resulting from chemical residual exposure. Cook distributed notices via courier on May 9, 2017. Customers were instructed to do the following: Distribution of the Cook Multi-Use Holmium Laser Fibers will not occur until the reprocessing instructions in the Instructions for Use have been corrected. Single use fibers are available. Contact your local sales representative for further information. You can continue to use your inventory of Cook Multi-Use Holmium Laser Fibers by following the attached Suggested Fiber Reprocessing Instructions. Action to be taken: 1. Examine your inventory immediately to identify and quarantine those affected products. 2. Implement the updated Suggested Fiber Reprocessing Instructions. 3. Return the required Acknowledgement and Receipt Form within 30 days. 4. Even if you do not have affected products, you must still complete the Acknowledgement and Receipt Form and send it via fax (812.339.7316) or email (fieldactionsna@cookmedical.com). 5. Immediately report any adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email at customerrelationsna@cookmedical.com. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: Online at: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail) Call the FDA at: 1-800-FDA-1088 Should you have any medical questions or concer
Initiated
2017-05-09
Root cause
Device Design
Terminated
2018-01-26
Distribution
Domestic: AL AR AZ CA CO FL GA HI ID IL IN KY LA MD MI MO MS NC NM NY OH OK OR PA SC TN TX UT VA WA WI Foreign: Austria, Australia, Argentina, Azerbaijan, Belgium, Bulgaria, Bahrain, Canada, Chile, China, Cyprus, Czech Republic, Denmark, Djibouti, Egypt, France, Hong Kong, Honduras, Croatia, Greece, Germany, Hungary, India, Ireland, Italy, Latvia, Malaysia, Morocco, Netherlands, Poland, Portugal, Puerto Rico, Saudi Arabia, Serbia, Spain, Slovenia, South Africa, Sri Lanka, Sweden, Singapore, Taiwan, Thailand, Tunisia, United Kingdom, and Uruguay
Recall number
Z-2539-2017
Recall status
Terminated
CDRH record ID
155666
FDA classified
2017-06-05
Recall event number
77304

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.