The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Source facts
Action
Customers were notified on 05/02/2018 via a MEDICAL DEVICE RECALL letter. Instructions included to immediately examine your inventory and quarantine any product subject to recall and replacement. Customers/distributors who received affected devices will be contacted to return their unopened/unused and/or reused stock of the affected product.
Initiated
2018-05-02
Root cause
Nonconforming Material/Component
Terminated
2019-02-08
Distribution
US Nationwide distribution in the states of CA, FL, IL, MI, NH, NY, OH, OR, PA, SC, UT, WV.
Worldwide distribution to Canada, Denmark, Germany, India, Israel, Japan, Russia, South Korea, Taiwan, United Kingdom.
Recall number
Z-2539-2018
Recall status
Terminated
CDRH record ID
165024
FDA record created
2018-07-25
Recall event number
80280