Open original FDA record ↗Recall record
- FDA identifier
- Z-2540-2016
- Company in FDA record
- TYRX Inc.
- FDA record creation date
- Aug 15, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
TRYX products are being recalled since the processes of spaying, welding, drying oven and polymer were not adequately validated.
Source facts
- Action
- Medtronic sent an Urgent Medical Device Recall letter dated June 2016 to affected customers. The letter identified the affected product problem and actions to be taken. Customers are asked to immediately remove and quarantine all unused product that remains in inventory, return unused product to Medtronic and contacted the local Medtronic representative or Customer Service at 800-848-9300 to assist with the return and credit of unused product. Medtronic will provide credit for all non-expired, unused product. The customer should complete the attached Customer Confirmation Certificate and email to RS.CFQFCA@Medtronic.com or fax to Medtronic at 651-367-0612 Attention Customer Focused Quality.
- Initiated
- 2016-06-29
- Root cause
- Process control
- Terminated
- 2017-12-12
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2540-2016
- Recall status
- Terminated
- CDRH record ID
- 148163
- FDA record created
- 2016-08-15
- Recall event number
- 74672