Open original FDA record ↗Recall record
- FDA identifier
- Z-2540-2019
- Company in FDA record
- Stryker Communications
- FDA record creation date
- Sep 12, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The back cover of the light head was insufficiently assembled and could potentially fall off the light.
Source facts
- Action
- The recalling firm issued letters via FedEx on 3/22/2019 to their U.S. consignees. The recall letter identified the affected product, problem and actions to be taken. ,The letter instructs the customer to disseminate it within their organization and that a Stryker representative will begin replacements for customer facilities in January. For the replacement, the Stryker representative will perform an onsite rework of the light head cover. In the meantime, the lights can continue to remain in full use. A response form was enclosed for the customer to indicate they have read and understood the instructions. The form was to be returned via email. The recalling firm's representatives will contact the customer to schedule an inspection of their light and will track completion of the visits and inspections.
- Initiated
- 2019-03-22
- Root cause
- Employee error
- Distribution
- Distribution was nationwide, including to Puerto Rico. There was also government/military distribution. Foreign distribution was made to Argentina, Austria, Australia, Belgium, Bulgaria, Canada, China, Colombia, Denmark, France, Germany, Hong Kong, India, Indonesia, Italy, Japan, Kazakhstan, Korea, Kuwait, Lithuania, Mexico, Netherlands, Philippines, Poland, Russia, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
- Recall number
- Z-2540-2019
- Recall status
- Open, Classified
- CDRH record ID
- 174897
- FDA record created
- 2019-09-12
- Recall event number
- 83355