Skip to content

Public FDA record

Z-2540-2020: FHC Guideline 4000 5.0 GL5 Main Processing Unit - Product Usage: intended to record and stimulate electr

Recall record Z-2540-2020 for FHC, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2540-2020
Company in FDA record
FHC, Inc.
FDA record creation date
Jul 8, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An electrical sub-circuit was not properly connected. Its purpose is to eliminate the possibility of interference with proper performance of the Guideline 5 Main Processing Unit, if an ESD event (greater than 6kV) were to occur.

Source facts

Action
FHC issued Urgent Medical Device Recall letter dated 6/17/20 via USPS certified return receipt,stating reason for recall, health risk and action to take: FHC, Inc. field representatives will be contacting you to return the effected product and issue a replacement. In the interim, we advise you to stop use of this product. Should you have any further questions regarding this matter you may contact me, by email at kmoeykens@fh-co.com or by telephone at 207-666-5425.
Initiated
2020-06-17
Root cause
Component change control
Terminated
2020-11-02
Distribution
US Nationwide distribution including in the states of CA, NC, TN, TX, WA.
Recall number
Z-2540-2020
Recall status
Terminated
CDRH record ID
182097
FDA record created
2020-07-08
Recall event number
85897

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗