Open original FDA record ↗Recall record
- FDA identifier
- Z-2540-2020
- Company in FDA record
- FHC, Inc.
- FDA record creation date
- Jul 8, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An electrical sub-circuit was not properly connected. Its purpose is to eliminate the possibility of interference with proper performance of the Guideline 5 Main Processing Unit, if an ESD event (greater than 6kV) were to occur.
Source facts
- Action
- FHC issued Urgent Medical Device Recall letter dated 6/17/20 via USPS certified return receipt,stating reason for recall, health risk and action to take: FHC, Inc. field representatives will be contacting you to return the effected product and issue a replacement. In the interim, we advise you to stop use of this product. Should you have any further questions regarding this matter you may contact me, by email at kmoeykens@fh-co.com or by telephone at 207-666-5425.
- Initiated
- 2020-06-17
- Root cause
- Component change control
- Terminated
- 2020-11-02
- Distribution
- US Nationwide distribution including in the states of CA, NC, TN, TX, WA.
- Recall number
- Z-2540-2020
- Recall status
- Terminated
- CDRH record ID
- 182097
- FDA record created
- 2020-07-08
- Recall event number
- 85897