Z-2540-2025: Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCMC use in conjunction with the Smart

Recall record Z-2540-2025 for Max Mobility LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2540-2025
Company in FDA record
Max Mobility LLC
FDA classification date
Sep 12, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.

Source facts

Action
On August 12, 2025, Permobil issued a "Urgent Medical Device Recall Expansion/Update" via email. This expansion supersedes the Class I recall ((Z-1116-2025, Z-1117-2025), initiated on December 20, 2024. Permobil asked consignees to take the following actions: 1. Immediately discontinue use of the SpeedControl dial to reduce the likelihood of a potentially hazardous or harmful situation. 1. Since all SpeedControl Dials are impacted by this recall, remove SpeedControl Dials from your sellable / distribution inventory 2. Use an alternate method to control the SmartDrive MX2+ Power Assist Device. Alternate controls include: " A wired SwitchControl, which can be ordered with a mono jack port to connect with an optional Buddy Button " Wirelessly through the SmartDrive MX2+ App utilizing a Apple iWatch or Samsung Galaxy watch. 3.Immediately contact any customers you have distributed the device to and provide a copy of this notice. It is also recommended that this notice be posted in your clinic/facility 4. You are required to acknowledge that you received this Urgent Medical Device Recall letter. Please complete within 10 days of receipt of the letter. Please use the QR code below to access the field action portal to record your acknowledgement
Initiated
2025-08-12
Root cause
Component design/selection
Distribution
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: CAN, NZL, KOR, BMU, GBR, JPN, TWN, AUS, BRA, COL, MEX, ABW, SWE, CHL, ARG, HKG
Recall number
Z-2540-2025
Recall status
Open, Classified
CDRH record ID
215324
FDA classified
2025-09-12
Recall event number
97356

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗