Z-2541-2015: CrossCHECK Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-L3008; 3.0 x 10 mm, REF CCP-L3010;

Recall record Z-2541-2015 for Wright Medical Technology, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2541-2015
Company in FDA record
Wright Medical Technology, Inc.
FDA classification date
Sep 1, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.

Source facts

Action
Wright sent an Urgent Medical Device Recall letter dated July 20, 2015, to all affected customers. Customers were asked to complete the attached response form confirming receipt of the notice and the status of any products on hand, and return a copy to Wright by email (complaints@wmt.com or fax (901)867-7401. Also, if customers have any product in their inventory the product should be sent back to Arlington, TN (ALM). For questions regarding this recall call 901-290-5662.
Initiated
2015-07-21
Root cause
Employee error
Terminated
2018-08-01
Distribution
Nationwide Distribution
Recall number
Z-2541-2015
Recall status
Terminated
CDRH record ID
139287
FDA classified
2015-09-01
Recall event number
71895

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗