Z-2541-2016: TRYX Neuro Absorbable Antibacterial Envelope Product Usage: Indicated for stabilization of implanted pac

Recall record Z-2541-2016 for TYRX Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2541-2016
Company in FDA record
TYRX Inc.
FDA record creation date
Aug 15, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

TRYX products are being recalled since the processes of spaying, welding, drying oven and polymer were not adequately validated.

Source facts

Action
Medtronic sent an Urgent Medical Device Recall letter dated June 2016 to affected customers. The letter identified the affected product problem and actions to be taken. Customers are asked to immediately remove and quarantine all unused product that remains in inventory, return unused product to Medtronic and contacted the local Medtronic representative or Customer Service at 800-848-9300 to assist with the return and credit of unused product. Medtronic will provide credit for all non-expired, unused product. The customer should complete the attached Customer Confirmation Certificate and email to RS.CFQFCA@Medtronic.com or fax to Medtronic at 651-367-0612 Attention Customer Focused Quality.
Initiated
2016-06-29
Root cause
Process control
Terminated
2017-12-12
Distribution
US Nationwide Distribution
Recall number
Z-2541-2016
Recall status
Terminated
CDRH record ID
148164
FDA record created
2016-08-15
Recall event number
74672

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗