Open original FDA record ↗Recall record
- FDA identifier
- Z-2541-2018
- Company in FDA record
- Bio-Rad Laboratories Inc
- FDA record creation date
- Jul 25, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system.
Source facts
- Action
- An Urgent Product Correction notification letter dated 4/2/18 was distributed to customers. The letter instructed customers to check the system and confirm it is labeled as shown in the letter. Then complete the attached form, indicating if the system includes a D-Link Brand USB hub, and fax or email back the form to Bio-Rad as soon as possible. If needed, a Bio-Rad Technical Support Representative will contact you shortly to arrange for replacement of the hub.
- Initiated
- 2018-04-02
- Root cause
- Device Design
- Distribution
- US Nationwide Distribution inn the states of : AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OR, PA, SC, TN, TX, VA, VT, WA, WI, and Puerto Rico.
- Recall number
- Z-2541-2018
- Recall status
- Terminated
- CDRH record ID
- 165373
- FDA record created
- 2018-07-25
- Recall event number
- 80403