Skip to content

Public FDA record

Z-2541-2019: Hemospray Endoscopic Hemostat

Recall record Z-2541-2019 for Wilson-Cook Medical Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2541-2019
Company in FDA record
Wilson-Cook Medical Inc.
FDA record creation date
Sep 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for Hemospray device to be unable to spray powder due to misassembly of devices.

Source facts

Action
Cook Medical notified customers starting on about 07/24/2019 via "URGENT: MEDICAL DEVICE RECALL" letter. Customers were informed that complaints have been received by Cook Medical that the Hemospray device is potentially unable to spray powder due to misassembly of the device. The potential risk involved includes delayed hemostasis, the need for additionaly hemostasis, and/or the patient may need to be transferred to surgery. If Hemostasis cannot be achieved in a timely manner in emergent cases, this may result in death. Instructions to customers include to immediately review the labels of impacted in inventory and quarantine any affected product that remains unused. All affected product on hand should be returned to Cook Medical and include the completed Acknowledgement and Receipt Form. The response form should be returned via fax (812.339.7316) or email FieldActionsNA@CookMedical.com. If the customer is a distributor, and the products have been further distributed, contact all customers and direct them to quarantine any affected product. For questions, please contact the Customer Relations Department at Cook Medical for more information or if you have questions. Please use FieldActionsNA@CookMedical.com, or call toll free 1-800-457-4500, press 4, then enter extension 15-2146, Monday through Friday, 8:00AM to 4:30PM, Eastern Time.
Initiated
2019-07-24
Root cause
Under Investigation by firm
Terminated
2020-06-15
Distribution
Nationwide distribution to CA, CO, CT, FL, IL, IN, KY, LA, MN, MO, NC, NJ, NY, OH, PA, RI, SD, TX, VA, WA. International distribution to Canada, Austria, Belgium, Cyprus, Denmark, France, Germany, Greece, Italy, Portugal, Spain, and United Kingdom.
Recall number
Z-2541-2019
Recall status
Terminated
CDRH record ID
175055
FDA record created
2019-09-13
Recall event number
83395

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.