Z-2542-2019: Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System, labeled as EMBLEMTM S-ICD Pulse Gener

Recall record Z-2542-2019 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2542-2019
Company in FDA record
Boston Scientific Corporation
FDA record creation date
Sep 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential need for device replacement - Elective Replacement Indicator (ERI) or End of Life (EOL) - earlier than expected due to compromised performance of an electrical component causing accelerated battery depletion.

Source facts

Action
Boston Scientific began notifying physicians of the recall on about 08/21/2019, via "URGENT: MEDICAL DEVICE ADVISORY" letter. The physicians associated with the affected devices were provide a letter with a list of affected model/serial numbers for patients which they have implanted and/or followed or sent to their center. The letter informed physicians of the potential for accelerated battery depletion on select active worldwide EMBLEM Subcutaneous Implantable Cardioverter Defibrillators (S-ICDs). Physicians were instructed to follow up with affected units implanted in patients by enroll and monitoring patients in LATITUDE to facilitate prompt detection of ERI/EOL during the interval between in-office device checks, perform a device follow-up every 3 months via remote or in-office interrogation, promptly investigate any suspected indication of accelerated depletion and contact Boston Scientific Technical Services for assistance as needed, and append the patients medical record with this letter to maintain awareness of this topic for the remaining service life of the device. Physicians were also instructed to evaluate the potential for life-threatening harm due to accelerated depletion for patients are the greatest risk, and replace the affected device within 21 days of ERI and prophylactically for those devices in high risk patients. Boston Scientific also provided an advisory letter for all EMBLEM physicians in order to reiterate and summarize the battery management recommendations that exist in current EMBLEM S-ICD labeling. Included in the customer notification to each physician was (affected and advisory) is a patient advisory letter, which each physician can chose at their discretion to provide. Questions and comments for US customers can be directed to Technical Services, Phone: 1.800.CARDIAC (227.3422) or Email: tech.services@bsci.com. In addition, Boston Scientific worked with the Heart Rhythm Society who sent an email August 22, 20
Initiated
2019-08-21
Root cause
Component design/selection
Distribution
Nationwide distribution to Alabama, Arizona, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Kansas, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Vermont, Virginia, Washington, Wisconsin, Wyoming. International distribution to Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Brazil, Brunei Darussalam, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Polynesia, Germany, Greece, Guam, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, , Lithuania, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Poland, Portugal, Puerto Rico, Qatar, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan R O C, Thailand, Tunisia, Turkey, U.S. - Virgin Islands, United Arab Emirates, United Kingdom.
Recall number
Z-2542-2019
Recall status
Open, Classified
CDRH record ID
175739
FDA record created
2019-09-13
Recall event number
83558

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗