Open original FDA record ↗Recall record
- FDA identifier
- Z-2543-2014
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Aug 22, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Persona Distal Valgus Alignment guide collet locks are seizing in the 'unlocked' position.
Source facts
- Action
- An Urgent Medical Device Correction letter dated, July 31, 2014, was issued to consignees. The letter identified the affected product, the problem, the risks, and responsibilities. Risk managers and surgeons are to review the notification and the updated surgical technique provided in Attachment 1; as well as, ensure that all affected personnel are aware of the contents. Questions or concerns should be directed to Zimmer customer call center at 1-877-946-2761.
- Initiated
- 2014-07-30
- Root cause
- Device Design
- Terminated
- 2015-03-06
- Distribution
- Worldwide Distribution -- United States, including Puerto Rico; and, to the countries of Australia, Austria, Japan, Belgium, Italy, Switzerland, India, Germany, Spain, France, Israel, Netherlands, Saudi Arabia, South Africa & Canada.
- Recall number
- Z-2543-2014
- Recall status
- Terminated
- CDRH record ID
- 128765
- FDA classified
- 2014-08-22
- Recall event number
- 68801