Skip to content

Public FDA record

Z-2543-2016: 1.5 mm HYDRO LeMaitre Valvulotome, 98 cm, Sterile REF # 1009-00 Product Usage: The HYDRO LeMaitre Valvu

Recall record Z-2543-2016 for LeMaitre Vascular, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2543-2016
Company in FDA record
LeMaitre Vascular, Inc.
FDA record creation date
Aug 16, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Wire which deploys the loop may be come detached at the handle causing the device to be inoperable

Source facts

Action
LeMaitre Vascular, Inc. sent an Urgent Medical Device Recall letter dated June 29, 2016 to customers via priority mail. The letter identified the affected product, problem and actions to be taken. Customers were requested to locate, quarantine and return the devices. Customers were instructed to complete and return the response form to LeMaitre Vascular, Inc. as a record of notification and reconciliation. Any unused product will be requested to be returned and replaced. For questions call 781-221-2266 ext. 183.
Initiated
2016-08-03
Root cause
Process control
Terminated
2021-04-08
Distribution
Worldwide Distribution - US Nationwide and in the countries of Australia, BELGIUM, BRAZIL, CANADA, Czech Republic, DENMARK, FINLAND, FRANCE, GERMANY, Italy, Japan, NETHERLANDS, Russia, Spain, SWEDEN, Switzerland, and UK.
Recall number
Z-2543-2016
Recall status
Terminated
CDRH record ID
148669
FDA record created
2016-08-16
Recall event number
74880

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗