Open original FDA record ↗Recall record
- FDA identifier
- Z-2544-2014
- Company in FDA record
- Medtronic MiniMed Inc.
- FDA classification date
- Aug 22, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.
Source facts
- Action
- The firm, Medtronic, sent an "IMPORTANT MEDICAL DEVICE SAFETY INFORMATION" letter dated March 13, 2014 to its customers. The letter described the product, problem and actions to be taken. The letter informs the customers that because accidental button pressing errors may occur it is important that customers always confirm the insulin dose flashing on the display is correct before pressing ACT to start delivery. The letter informs the customers of what safety settings will minimize the risk of an accidental programming error and what actions to be taken. Should you have any questions about this letter or concerns with your keypad, please do not hesitate to contact the 24 hour helpline at (866) 222-7304 for a replacement.
- Initiated
- 2014-03-13
- Root cause
- Device Design
- Terminated
- 2015-04-07
- Distribution
- Worldwide Distribution-Nationwide in US
- Recall number
- Z-2544-2014
- Recall status
- Terminated
- CDRH record ID
- 127259
- FDA classified
- 2014-08-22
- Recall event number
- 68277