Skip to content

Public FDA record

Z-2544-2020: Cardinal Health NPWT Occlusion Detection Canister 300cc, Prescription Only - Product Usage: The systems a

Recall record Z-2544-2020 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2544-2020
Company in FDA record
Cardinal Health 200, LLC
FDA record creation date
Jul 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Canisters potentially contain an oversized O-ring that cannot be installed into the mating connector on the device.

Source facts

Action
Cardinal Health issued a letter titled, "Urgent Medical Device Recall", to customers on 06/04/2020. The letter advises customers of a voluntarily recalling associated with lots of Cardinal Health NPWT 300cc Canister with gel and Cardinal Health NPWT 300cc Canister with gel Occlusion Detection. The recall is being conducted due to Canisters potentially containing an oversized O-ring. The canisters with the oversized O-ring cannot be installed into the mating connector on the device causing the device to be unusable. According to the letter, customers should check all storage and usage locations to confirm whether they have any units of the affected product, quarantine all on-hand product and return the enclosed acknowledgment form via fax to 847-689-9101 or 614-652-9648 or email at GMB-FieldCorrectiveAction@cardinalhealth.com. The returned form should indicate the product code, lot and quantity of product quarantined or discarded. Response is needed whether or not customers have affected product. Customers should notify any customers to whom they may have distributed, or forward the product affected by the recall and should include a copy of the recall notification letter. The firm customer service is available Monday to Friday between 8:00am - 5:00pm Eastern Time. The firm contact information is listed as: Hospital: 800-964-5227, Federal Government: 800-444-1166, Distributor: 800-635-6021, All other Customers: 888-444-5440.
Initiated
2020-06-04
Root cause
Nonconforming Material/Component
Terminated
2025-08-05
Distribution
Worldwide distribution - US Nationwide distribution and the country of Canada.
Recall number
Z-2544-2020
Recall status
Terminated
CDRH record ID
181831
FDA record created
2020-07-10
Recall event number
85803

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗