Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from
each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
Source facts
Action
Zimmer sent an URGENT MEDICAL DEVICE RECALL letter dated July 2014 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Please call the Customer Call Center at 1-800-348-2759 for questions or concerns involving this notice.
Initiated
2014-07-22
Root cause
Package design/selection
Terminated
2015-09-15
Distribution
Worldwide Distribution.
Recall number
Z-2546-2014
Recall status
Terminated
CDRH record ID
128787
FDA classified
2014-08-26
Recall event number
68818