Multi-Function Defibrillation Electrodes, Pre connect-
potential for damage to the wire insulation may
result in a potentially increased risk for reduced or no patient therapy and patient and/or clinician burns
Source facts
Action
Medtonic/Covidien issued recall letter on August 19, 2015. Customers were notified via Federal Express and the letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 2 Ludlow Park Drive, Chicopee, MA 01022, Attention: Field Action Returns.
Contact your Medtronic representative or Customer Service at (800)-882-5878.
Initiated
2015-08-19
Root cause
Process control
Terminated
2017-05-24
Distribution
Nationwide
Foreign:
Canada
Australia
Denmark
Germany
New Zealand
Recall number
Z-2546-2015
Recall status
Terminated
CDRH record ID
139462
FDA classified
2015-09-02
Recall event number
71963