Z-2548-2026: Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscill

Recall record Z-2548-2026 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2548-2026
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Jul 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

Source facts

Action
Baxter issued an URGENT MEDICAL DEVICE CORRECITON notice to its consignees on 5/21/2026 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: "Actions to be Taken by Customers 1. Before each use, carefully inspect the nebulizer cup for damage and ensure it is fully and securely locked during assembly, following the enclosed instructions. If leakage is observed during therapy, discontinue the use of that nebulizer and replace the patient circuit. 2. Nebulizer cups with patient circuits are routinely replaced as part of normal therapy usage. Additional patient circuits can be provided through standard distribution. 3. If you received this communication directly from Baxter, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any remaining inventory. Log in to the portal using the account number listed in the enclosed reply instruction sheet. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this urgent medical device correction in accordance with your customary procedures and check the associated box
Initiated
2026-05-21
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall number
Z-2548-2026
Recall status
Open, Classified
CDRH record ID
220862
FDA classified
2026-07-01
Recall event number
99119

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗