Skip to content

Public FDA record

Z-2549-2019: Philips HeartStart Infant/Child SMART Pads Cartridge, Model # M5072A, UDI:00884838023758

Recall record Z-2549-2019 for Philips North America LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2549-2019
Company in FDA record
Philips North America LLC
FDA record creation date
Sep 20, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The recall is being initiated due to incorrect age range labeling (8+ yrs) on the foil packaging for infant/child pad cartridges. The infant/child pad cartridge is designed for use with the defibrillator An Infant/Child pad cartridge should not be used on an adult because it delivers lower energy (50 J vs 150 J) than the Adult pads cartridge and may therefore not be effective on an adult.

Source facts

Action
A labeling specification for the HS1 is available against which labeling can be checked prior to design transfer and/or localization. This activity has been moved under regulatory affairs for review and maintenance.
Initiated
2018-08-08
Root cause
Labeling False and Misleading
Terminated
2021-04-29
Distribution
US: AK, AL,AR,AZ,CA,CO,CT,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV, WY, Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Denmark, Dominican Republic, Egypt, El Salvador, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Republic of, Malaysia, Malta, Martinique, Mauritius, Mexico, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Norway, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, R¿union, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Viet Nam,
Recall number
Z-2549-2019
Recall status
Terminated
CDRH record ID
176133
FDA record created
2019-09-20
Recall event number
83681

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗