Open original FDA record ↗Recall record
- FDA identifier
- Z-2550-2016
- Company in FDA record
- Resmed Corporation
- FDA record creation date
- Aug 16, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Masks in this lot, labelled and sold as 62702 vented Quattro Air FFM size Medium actually contain a non-vented (NV) Quattro Air FFM Large mask.
Source facts
- Action
- An urgent product recall notification letter was sent to all customers on 7/12/16 to inform them that ResMed is recalling NV Quattro Air & Quattro Air Mask due to a mislabeling Issue. The letter informs the customers that during the investigation of a product complaint, ResMed identified that a small number of masks labelled and sold as 62702 Quattro Air FFM MED AMER, Lot 1142623 actually contain an NV Quattro Air FFM Large mask. The letter informs the customers to notify anyone they may have sold the products to. Customers with any questions are instructed to call the ResMed Customer Service team: 858-836-5968.
- Initiated
- 2016-07-12
- Root cause
- Packaging process control
- Terminated
- 2016-09-26
- Distribution
- US nationwide distribution
- Recall number
- Z-2550-2016
- Recall status
- Terminated
- CDRH record ID
- 148349
- FDA record created
- 2016-08-16
- Recall event number
- 74736