Open original FDA record ↗Recall record
- FDA identifier
- Z-2551-2020
- Company in FDA record
- Roche Molecular Systems, Inc.
- FDA record creation date
- Jul 10, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.
Source facts
- Action
- On June 5, 2020, the firm distributed Urgent Medical Device Correction letters to consignee accounts. The UMDC instructs the customers to determine whether or not their laboratory is using the affected Sample Transfer protocol version 3.0 with their MagNA Pure 96 instrument. If the affected protocol is found on the instruments Control Unit, it needs to be immediately deleted by following the instructions provided in the UMDC. A newly updated Sample Transfer protocol version 4.0 is available. Roche Affiliate organizations will provide instructions for obtaining the new version of the protocol. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-1247 if you have questions about this recall.
- Initiated
- 2020-06-03
- Root cause
- Software change control
- Terminated
- 2023-10-04
- Distribution
- Worldwide distribution - US Nationwide distribution.
- Recall number
- Z-2551-2020
- Recall status
- Terminated
- CDRH record ID
- 181876
- FDA record created
- 2020-07-10
- Recall event number
- 85838