Open original FDA record ↗Recall record
- FDA identifier
- Z-2552-2016
- Company in FDA record
- Reshape Medical Inc
- FDA record creation date
- Aug 17, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a suspected misalignment of the connection between the proximal balloon cap of the implant and the delivery catheter. The possible misalignment is the result of an assembly error during manufacturing and may lead to leakage of saline solution during the balloon inflation procedure.
Source facts
- Action
- A Dear Doctor letter dated May 4, 2016, was sent to all the customers to inform them that ReShape Medical is recalling ReShape Integrated Dual Balloon System, Gen 1 due to a suspected misalignment of the connection between the proximal balloon cap and the delivery catheter. The letter informs the customers that the possible misalignment is the result of an assembly error during manufacturing and may lead to leakage of saline solution during the balloon inflation procedure. Customers are informed of the planned field correction and if customers have any questions then they're instructed to contact ReShape Medical Customer Service at (844) 937-7374. For further questions please call (949) 429-6680 ext. 106
- Initiated
- 2016-05-04
- Root cause
- Process control
- Terminated
- 2016-08-17
- Distribution
- US: Distribution to the states of : NY, OH and CA.
- Recall number
- Z-2552-2016
- Recall status
- Terminated
- CDRH record ID
- 148437
- FDA record created
- 2016-08-17
- Recall event number
- 74781