Skip to content

Public FDA record

Z-2552-2019: Alinity c Lithium Reagent Kit- intended for the quantitation of lithium in serum or plasma using respecti

Recall record Z-2552-2019 for Sentinel CH SpA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2552-2019
Company in FDA record
Sentinel CH SpA
FDA record creation date
Sep 20, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate Dehydrogenase (LDH) reagent (LN 2P56) may result infalsely elevated lithium patient results

Source facts

Action
Sentinel issued Urgent Field Safety Notice letter on 6/28/19 via email to the distributor. The Distributor Abbott notified customers by letter Urgent Field Safety Notice Product Correction dated July 11, 2019 advising of the problem, health risk and action to take: If the LDH reagent is installed on your instrument, please follow the instructions 1. Update the Lithium (LN 8L25-30) assay parameters to add the new SmartWash. For additional information, see the ARCHITECT System Operations Manual, Section 2, Configure the SmartWash settings (c System. Install the Alinit y c Lithium assay file version 6 to implement the new LDH SmartWash. Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service
Initiated
2019-06-28
Root cause
Under Investigation by firm
Terminated
2020-07-24
Distribution
Nationwide Foreign: ARGENTINA AUSTRALIA AUSTRIA AZERBAIJAN BANGLADESH BELARUS BELGIUM BOSNIA AND HERZEGOVI BRAZIL BRUNEI CANADA CHILE COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK FAROE ISLANDS FINLAND FRANCE GERMANY GREECE GREENLAND HONG KONG IRELAND ISRAEL ITALY JORDAN KUWAIT LEBANON LUXEMBOURG MALAYSIA MALDIVES MEXICO MOROCCO NETHERLANDS NEW ZEALAND NIGERIA NORWAY OMAN PAKISTAN PANAMA PARAGUAY PHILIPPINES POLAND PORTUGAL PUERTO RICO ROMANIA RWANDA SAUDI ARABIA SLOVAKIA SOUTH AFRICA SPAIN ST. VINCENT SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UAE UNITED KINGDOM URUGUAY
Recall number
Z-2552-2019
Recall status
Terminated
CDRH record ID
175597
FDA record created
2019-09-20
Recall event number
83516

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗