Open original FDA record ↗Recall record
- FDA identifier
- Z-2552-2020
- Company in FDA record
- Bausch & Lomb Surgical, Inc.
- FDA record creation date
- Jul 10, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The toric axis marks were misaligned (from the expected position in relation to the haptics) on an implanted lens.
Source facts
- Action
- On 06/09/2020, BAUSCH and LOMB's Sales Representatives contacted customers by telephone and email. In addition, the firm sent a letter to customers dated, June 17, 2020 and titled, "MEDICAL DEVICE VOLUNTARY PRODUCT RECALL". The letter informs customers about a voluntary recall of one lot of enhanced enVista MX60ET intraocular lenses (IOLs). Customers should quarantine product and return product to the firm. Customers should complete the enclosed Medical Device Voluntary Recall Acknowledgement Form and contact the firm to obtain a Return Material Authorization Number and arrange for pickup of the identified product. The contact information for Bausch and Lomb Surgical Customer Service is 1-800-338-2020, Customer Service (option 2).
- Initiated
- 2020-06-09
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2021-02-11
- Distribution
- U.S. Nationwide distribution including in the states of IN, SC, CT, NV, and NC.
- Recall number
- Z-2552-2020
- Recall status
- Terminated
- CDRH record ID
- 181955
- FDA record created
- 2020-07-10
- Recall event number
- 85862