The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator.
Source facts
Action
Siemens notified end users of a potential safety issue with the Reagent Lid hinge on the Dimension EXL instrument via an Urgent Medical Device Correction letter, dated July 2014. The letter indicated Customer Service Engineers will inspect and adjust the tension on the reagent lid hinges during every visit and replace hinges which no longer properly adjust. Users were urged to use caution when the Reagent Lid is open and notify Siemens the Reagent Lid is shifting down.
Initiated
2014-07-16
Root cause
Device Design
Terminated
2014-12-16
Distribution
Nationwide and internationally, including Canada and Mexico
Recall number
Z-2553-2014
Recall status
Terminated
CDRH record ID
128863
FDA classified
2014-08-26
Recall event number
68854