Open original FDA record ↗Recall record
- FDA identifier
- Z-2553-2025
- Company in FDA record
- Merit Medical Systems, Inc.
- FDA classification date
- Sep 11, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a manufacturing issue involving the silicone quad ring where the ring is deformed or otherwise damaged and may cause a foreign body in the fluid pathway.
Source facts
- Action
- On 07/23/2025, the firm emailed an "URGENT MEDICAL DEVICE RECALL NOTICE" to customers informing them that due to a manufacturing defect. Specifically, the o-ring between the y-body and the rotator may be damaged during assembly and a portion of the o-ring may remain loose or detach. Use of the affected product may result in a foreign body in the fluid path. Customer are instructed to: Immediately stop using or distributing the affected lots and return to Merit. 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Please fill out, scan and email the completed Customer Response Form to Customer Service at response@merit.com within 10 business days. All affected product shipped to you must be accounted for on the CRF. 5. Please immediately return all affected lots in your possession to Merit, per the instructions in the attached CRF. Any questions concerning this communication, contact your Merit Sales Representative or Merit Customer Service via email at RESPONSE@merit.com or via phone at +1 800 356 3748 | Hours: 6 am to 6 pm MST | Mon-Fri.
- Initiated
- 2025-07-25
- Root cause
- Employee error
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, FL, LA, MT, NY, SD, and WI. The countries of Canada and Korea
- Recall number
- Z-2553-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215547
- FDA classified
- 2025-09-11
- Recall event number
- 97420