Terumo has discovered several instances of adhesive build-up and or skiving of plastic (plastic flash) inside the distal tip of the Terumo Coronary Ostia Cannula. The defect was discovered during in-house inspection after Terumo CVS implemented a new inspection process for the cannula. There have been no reports of the defect in clinical use.
Source facts
Action
Terumo Cardiovascular Systems Corporation sent an URGENT MEDICAL DEVICE REMOVAL notice dated May 16, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken. The letter described the problem, the potiential hazard, and the action to be taken by the consignee. Consignees were instructed to discontinue use of the product and return all product to Terumo. The letter provided a list of alternate products to be used instead of the recalled products. Consignees with questions should call 1-800-521-2818 Monday through Friday 8AM to 6PM.
Initiated
2011-05-16
Root cause
Process control
Terminated
2012-03-21
Distribution
Worldwide Distribution - USA (nationwide) and the countries of BELGIUM, BRAZIL, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES (UAE), SINGAPORE, MALAYSIA, AND SOUTH AFRICA.
Recall number
Z-2554-2011
Recall status
Terminated
CDRH record ID
100266
FDA classified
2011-06-14
Recall event number
58762