The firm became aware of a potential problem that was initiated by a customer complaint. After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor. This change to the resistor was a planned change to address a product improvement (improve battery drain). However, this modification rendered the monitor's alarm via the nurse call communication port to not function.
Source facts
Action
Urgent Medical Device Recall Letters (dated 8/01/14) and Response Forms were sent to the consignees via e-mail (w/read receipt) on 8/05/14.
Initiated
2014-08-05
Root cause
Component design/selection
Terminated
2017-02-28
Distribution
Nationwide, Canada, Ireland, Italy
Recall number
Z-2554-2014
Recall status
Terminated
CDRH record ID
129118
FDA classified
2014-08-26
Recall event number
68961