Z-2554-2018: RayStation Radiation Therapy Treatment Planning System, Model nos. 2.5, 3.5, 4.0, 4.3, 4.5, 4.7, 4.9, 5.0

Recall record Z-2554-2018 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2554-2018
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA record creation date
Jul 26, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has learned that some RayStation/RayPlan users have commissioned machines with erroneous Beam profile correction parameters. These parameters affect the dose calculated in corners of large or off-axis fields. This effect cannot be seen in the Beam Commissioning module and dose in large or off-axis fields needs to be validated using the Beam 3D Modeling module in RayPhysics/RayPlan Physics. The user must be aware to avoid incorrect dose calculations during treatment planning.

Source facts

Action
On June 6, 2018, the firm issued an Urgent Field Safety Notice, Medical Device Correction letter to affected customers via email. The notice was a reminder to follow the instructions listed in the product documentation when creating a photon beam model in RayStation/RayPlan. ACTIONS TO BE TAKEN BY THE USER " Always validate the beam model for all clinically relevant fields, including, but not limited to, large and off-axis fields, before the system is used to create clinical treatment plans. " Carefully review all beam model parameters before commissioning. " Be aware that the Beam profile correction values may need to be adjusted in order to correctly model the primary collimator that limits the dose in the corners of large or off-axis fields. " Review any existing photon beam models to ensure that the actions above have been properly performed. Please educate physics staff about these user responsibilities. Inspect your product and identify all installed units with the above software version number(s), then confirm that you have read and understood this notice by replying to the notification email. Safe use relies on careful validation of the beam model for all clinically relevant fields before the system is used to create clinical treatment plans. This information is already present in the Instructions For Use, the Reference Manual and the RayPhysics/RayPlan Physics Manual for the existing product versions. To further reduce the risk of using improper beam models, the safety warnings within the product documentation will be updated in the next version of RayStation/RayPlan, scheduled for market release in June 2018.
Initiated
2018-06-06
Root cause
Software design
Terminated
2019-04-04
Distribution
US Nationwide distribution.
Recall number
Z-2554-2018
Recall status
Terminated
CDRH record ID
165653
FDA record created
2018-07-26
Recall event number
80515

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗