Open original FDA record ↗Recall record
- FDA identifier
- Z-2554-2018
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA record creation date
- Jul 26, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has learned that some RayStation/RayPlan users have commissioned machines with erroneous Beam profile correction parameters. These parameters affect the dose calculated in corners of large or off-axis fields. This effect cannot be seen in the Beam Commissioning module and dose in large or off-axis fields needs to be validated using the Beam 3D Modeling module in RayPhysics/RayPlan Physics. The user must be aware to avoid incorrect dose calculations during treatment planning.
Source facts
- Action
- On June 6, 2018, the firm issued an Urgent Field Safety Notice, Medical Device Correction letter to affected customers via email. The notice was a reminder to follow the instructions listed in the product documentation when creating a photon beam model in RayStation/RayPlan. ACTIONS TO BE TAKEN BY THE USER " Always validate the beam model for all clinically relevant fields, including, but not limited to, large and off-axis fields, before the system is used to create clinical treatment plans. " Carefully review all beam model parameters before commissioning. " Be aware that the Beam profile correction values may need to be adjusted in order to correctly model the primary collimator that limits the dose in the corners of large or off-axis fields. " Review any existing photon beam models to ensure that the actions above have been properly performed. Please educate physics staff about these user responsibilities. Inspect your product and identify all installed units with the above software version number(s), then confirm that you have read and understood this notice by replying to the notification email. Safe use relies on careful validation of the beam model for all clinically relevant fields before the system is used to create clinical treatment plans. This information is already present in the Instructions For Use, the Reference Manual and the RayPhysics/RayPlan Physics Manual for the existing product versions. To further reduce the risk of using improper beam models, the safety warnings within the product documentation will be updated in the next version of RayStation/RayPlan, scheduled for market release in June 2018.
- Initiated
- 2018-06-06
- Root cause
- Software design
- Terminated
- 2019-04-04
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2554-2018
- Recall status
- Terminated
- CDRH record ID
- 165653
- FDA record created
- 2018-07-26
- Recall event number
- 80515