Reason stated in the source
Customers who utilize PharmGuard Server Software (PGS) with Medfusion 4000 pumps manufactured prior to November 10, 2017, may experience wireless network connectivity issues on certain pumps due to a potential Radio Module issue on some Medfusion 4000 pumps.
Source facts
- Action
- Smiths Medical notified customers on about 10/15/2018 via "URGENT MEDICAL DEVICE FIELD SAFETY NOTICE". Specifically, Biomed individuals (owners) at each facility were notified because they are in contact with Smiths Medical service and repair team to conduct the correction. Instructions included the following actions: 1. To determine if your facility has any devices with a duplicate wireless MAC Address, a software script must be run on the PGS Database to collect MAC Address and serial number information. The PGS Database must be at version 2.0, 2.1, or 2.2 for this data collection process to be completed. Refer to instructions on Attachment 1, Field Safety Notice Response Form, for available options to complete this data collection process, and if needed, to upgrade your PGS database and Medfusion 4000 device software/firmware. Please return the Response Form within 10 days of receipt of the Field Safety Notice. 2. If it is determined that any of your Medfusion 4000 pumps are affected by this issue, Smiths Medical Service and Repair will contact you to discuss repair options for the affected devices, based on the contact information provided on the Response Form. Questions regarding the notification can be sent to Smiths Medical via email at macaddress@smiths-medical.com.
- Initiated
- 2018-10-15
- Root cause
- Process control
- Terminated
- 2022-10-11
- Distribution
- Worldwide distribution - Nationwide in the states of AL, AR, AZ, CA, CN, CO, CT, DC, FL, GA, ID, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV and country of Canada.
- Recall number
- Z-2554-2019
- Recall status
- Terminated
- CDRH record ID
- 168030
- FDA record created
- 2019-09-23
- Recall event number
- 81283