Z-2554-2019: Medfusion Model 4000 Syringe Infusion Pump, with PharmGuard Server Software (PGS). Sold under the foll

Recall record Z-2554-2019 for Smiths Medical ASD Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2554-2019
Company in FDA record
Smiths Medical ASD Inc.
FDA record creation date
Sep 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customers who utilize PharmGuard Server Software (PGS) with Medfusion 4000 pumps manufactured prior to November 10, 2017, may experience wireless network connectivity issues on certain pumps due to a potential Radio Module issue on some Medfusion 4000 pumps.

Source facts

Action
Smiths Medical notified customers on about 10/15/2018 via "URGENT MEDICAL DEVICE FIELD SAFETY NOTICE". Specifically, Biomed individuals (owners) at each facility were notified because they are in contact with Smiths Medical service and repair team to conduct the correction. Instructions included the following actions: 1. To determine if your facility has any devices with a duplicate wireless MAC Address, a software script must be run on the PGS Database to collect MAC Address and serial number information. The PGS Database must be at version 2.0, 2.1, or 2.2 for this data collection process to be completed. Refer to instructions on Attachment 1, Field Safety Notice Response Form, for available options to complete this data collection process, and if needed, to upgrade your PGS database and Medfusion 4000 device software/firmware. Please return the Response Form within 10 days of receipt of the Field Safety Notice. 2. If it is determined that any of your Medfusion 4000 pumps are affected by this issue, Smiths Medical Service and Repair will contact you to discuss repair options for the affected devices, based on the contact information provided on the Response Form. Questions regarding the notification can be sent to Smiths Medical via email at macaddress@smiths-medical.com.
Initiated
2018-10-15
Root cause
Process control
Terminated
2022-10-11
Distribution
Worldwide distribution - Nationwide in the states of AL, AR, AZ, CA, CN, CO, CT, DC, FL, GA, ID, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV and country of Canada.
Recall number
Z-2554-2019
Recall status
Terminated
CDRH record ID
168030
FDA record created
2019-09-23
Recall event number
81283

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗