Z-2555-2014: The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, and d

Recall record Z-2555-2014 for Thermedx LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2555-2014
Company in FDA record
Thermedx LLC
FDA classification date
Aug 27, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The deficit display may be higher than the actual deficit when the fluid collection canister is fully loaded.

Source facts

Action
Thermedx sent an Urgent Product Correction letter dated March 15, 2014, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were notified that the devices were in need of correction, and that it would be performed at their facilities by Thermedx. Customers with questions were instructed to call a Thermedx technical service representative at 1-888-542-9276. For questions regarding this recall call 440-542-0883.
Initiated
2014-02-10
Root cause
Component design/selection
Terminated
2015-08-04
Distribution
Nationwide Distribution including HI, MS, NY, OH, PA, VA, WA, WV, and IL.
Recall number
Z-2555-2014
Recall status
Terminated
CDRH record ID
129203
FDA classified
2014-08-27
Recall event number
68986

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗