Reason stated in the source
Possible failure of screws which hold the gas spring fixation bracket in place. Due to this failure, the gas spring, which is fixed to the upper part of the parallelogram, cannot hold up the weight of the optics carrier.
Source facts
- Action
- Leica Microsystems notified customers on about 06/04/2018 via URGENT FIELD SAFETY NOTICE letter. The letter informed customers that they would be contacted by a Leica Microsystems representative to arrange for the field update in the next 12 months. Customers were instructed that they could continue to use the system pending the update if they ensure a functional check is performed prior to surgery and follow user manual directions. If any anomalies are noticed prior to the correction, customers should contact their Leica representative. Customers were also instructed to provide the notification to all appropriate personnel , notify customers if the unit was transferred, and complete and return the Medical Device Correction Form. A follow-up recall notification was sent to customers on or about 06/07/2018. The follow up letter discusses the functional check prior to use and provides additional compliant information.
- Initiated
- 2018-06-04
- Root cause
- Nonconforming Material/Component
- Terminated
- 2021-03-25
- Distribution
- Worldwide Distribution US Nationwide in the states of : CA, CO, CT, FL, HI, ID, IL, KS, KY, LA, MD, MI, MN, MO, NC, NJ, NY, OH, PA, SD, TN, TX. International Distribution: Canada, AUSTRALIA, BELGIUM, Bolivia, BRAZIL, BULGARIA, CHILE, CHINA, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, HONGKONG, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KUWAIT, MALAYSIA, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, TAIWAN,THAILAND, TURKEY, UK
- Recall number
- Z-2555-2018
- Recall status
- Terminated
- CDRH record ID
- 164913
- FDA record created
- 2018-07-26
- Recall event number
- 80229