- FDA identifier
- Z-2555-2025
- Company in FDA record
- Orthofix U.S. LLC
- FDA classification date
- Sep 11, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A 37 mm implant width intervertebral body fusion device has a labeling discrepancy with incorrect width of 33 mm stated on the outer label side flap and inner label side flap information box. The 37 mm width on the front labels is correct.
Source facts
- Action
- On 6/24/2025, recall notes were emailed to customers who were asked to do the following: 1) Examine your inventory and quarantine any affected product. 2) If implanted, healthcare Providers should determine whether it is necessary to inform the patient, based on clinical judgment and individual circumstances. 3) This notice should be shared with all relevant personnel within your organization and forwarded to any other organization or facility where potentially affected devices may have been transferred. 4) Complete and return the response form to the firm. Should you have any questions regarding this matter, please contact the firm at madison.ebel@seaspine.com
- Initiated
- 2025-06-24
- Root cause
- Labeling design
- Distribution
- US Nationwide distribution in the states of CO, CA.
- Recall number
- Z-2555-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215012
- FDA classified
- 2025-09-11
- Recall event number
- 97265