Open original FDA record ↗Recall record
- FDA identifier
- Z-2558-2011
- Company in FDA record
- OPTI Medical Systems, Inc
- FDA classification date
- Jun 14, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The values reported by the chloride sensors may report erroneous values below specification on aqueous controls and potentially with patient samples of whole blood and serum/plasma.
Source facts
- Action
- OPTIMedical Systems sent an "URGENT: MEDICAL DEVICE RECALL" letter dated March 15, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were notified by letter via e-mail and fax. Customers were instructed to discontinue use of the product and to return any unused product to OPTI Medical Systems, Inc. Customers outside the U.S. were instructed to destroy unused product and certify the destruction on the enclosed Recall Response Card. The Response Card was to be completed and returned to OPTI Medical Customer Service via fax. OPTI Medical Systems will replace the cassettes they have with product from another lot. Customers were instructed to contact the Technical Support Department at 1-800-490-6784 for questions relating to this notice.
- Initiated
- 2011-03-15
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-08-08
- Distribution
- Worldwide Distribution-- USA (nationwide) including states of CA and VA, and countries of India, Lebanon, Philippines, Sweden, Venezuela and Vietnam.
- Recall number
- Z-2558-2011
- Recall status
- Terminated
- CDRH record ID
- 98847
- FDA classified
- 2011-06-14
- Recall event number
- 58272