Instrument stopped working due to a software lockup, and no patient results are produced.
Source facts
Action
In very close cooperation with the consignees, the affected end users running on EPROM version V1.2.3 were contacted. Initially, a temporary work around (reset of the software) was communicated with the consignees. The consignees devices were installed to document all errors during runtime .In addition, a new software was developed and tested. The new software resulted in version 1.2.5 of EPROM, which was released June 23, 2016. Consignees of the V-Twin all received a field update kit with the new software.
Initiated
2016-01-18
Root cause
Software Manufacturing/Software Deployment
Distribution
US in the state of New York.
Recall number
Z-2558-2019
Recall status
Open, Classified
CDRH record ID
175935
FDA record created
2019-09-23
Recall event number
83607