Open original FDA record ↗Recall record
- FDA identifier
- Z-2560-2011
- Company in FDA record
- Computerized Medical Systems Inc
- FDA classification date
- Jun 14, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Monaco: If the user selects Monte Carlo Photon as the Secondary Algorithm (to be used for final dose calculation through segments), and selects SKIP Forward button at any time during stage 2 optimization, the system may only use the pencil beam algorithm for final computation of dose even though Monte Carlo was selected. Other than specific output to the console window, there will be no other in
Source facts
- Action
- Computerized Medical Systems, Inc. sent a "MONACO CUSTOMER ADVISORY" letter dated June 2008 to all affected customers. The letter describes the product, problem, and the action to be taken. The customer was instructed to use caution when using the Skip Forward button in stage 2 according to written instructions provided in the letter. This defect exists in Monaco Release 1.0.0 through 1.0.2 and will be resolved as soon as possible. Note: The issue was resolved in Monaco Release 2.0.1 and above. Questions should be directed to the manufacturer's representative telephone number at 408-830-8023.
- Initiated
- 2008-06-30
- Root cause
- Software design
- Terminated
- 2013-04-04
- Distribution
- Worldwide Distribution-- USA (nationwide) including the states of IL, MI, MO, OR, SC, TX, VA, and Puerto Rico and the countries of Canada, China, Cyprus, Estonia, France, Germany, Greece, Italy, Japan, Netherlands, Philippines, Poland, Portugal, Slovenia, South Korea, Spain, and United Kingdom.
- Recall number
- Z-2560-2011
- Recall status
- Terminated
- CDRH record ID
- 98760
- FDA classified
- 2011-06-14
- Recall event number
- 58229