Weld fractures on the Sleeve Installation Tool for ACL reconstruction system.
Source facts
Action
OrthoPediatrics sent a Notification letter dated July 25, 2014 to all affected customers. The letter Identified the affected product, problem and actions to be taken. The letter informed customers to take immediate possession of all affected product within their control and at any of their health care facilities and immediately return the recalled device(s) to OrthoPediatrics Corp. This notice and the attached Reply Form should be forwarded if the affected product was further distributed. .
Initiated
2014-07-25
Root cause
Nonconforming Material/Component
Terminated
2015-05-19
Distribution
US Nationwide Distribution.
Recall number
Z-2560-2014
Recall status
Terminated
CDRH record ID
129442
FDA classified
2014-08-29
Recall event number
69050