Public FDA record

Z-2560-2018: 1) Drainer Centesis Catheter, Sterile EO, Rx Only, Catalog #: a) DRC-001-05, b) DRC-001-06, c) DRC-001-1

Recall record Z-2560-2018 for Galt Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2560-2018
Company in FDA record
Galt Medical Corporation
FDA record creation date
Jul 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.

Source facts

Action
On approximately 05/01/2018 recall letters and customer response forms were mailed to all consignees.
Initiated
2018-05-02
Root cause
Other
Terminated
2020-07-20
Distribution
U.S.: MN, PA, VA, CA, UT, FL, NY, GA, LA, MO, TX, MS, OH, WA, MA, IL, KY, NC, and Puerto Rico; OUS: Canada, Australia, France, Germany, Ireland, Italy, Netherlands, Poland, Slovakia, Switzerland, Belgium, United Kingdom, Korea
Recall number
Z-2560-2018
Recall status
Terminated
CDRH record ID
164284
FDA record created
2018-07-27
Recall event number
80021

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗