Z-2560-2025: VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18

Recall record Z-2560-2025 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2560-2025
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Sep 12, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Source facts

Action
On 29 July 2025, QuidelOrtho mailed a hardcopy customer letter (Ref. CL2025-201) via FedEx overnight courier and/ or electronically sent the customer letter using ORTHO PLUS e-Communications to customers. Consignees were instructed to load ADD DRV 6339 (or above) on your VITROS XT 3400 and/or VITROS XT 7600 System(s), complete and return the provided Confirmation of Receipt, save the notification with your User Documentation or post this notification by each VITROS XT 3400/VITROS XT 7600 System in your laboratory until the issue has been resolved, and forward the information if affected units were further distributed.
Initiated
2025-07-29
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-2560-2025
Recall status
Open, Classified
CDRH record ID
215636
FDA classified
2025-09-12
Recall event number
97409

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗