It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.
Source facts
Action
The firm notified their consignees of the issue by email on 09/04/2019. The letter explained the problem and the workaround and requested the notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed and that all appropriate personnel working with the product be advised of the content of the letter.
Initiated
2019-09-04
Root cause
Software design
Distribution
Worldwide Distribution - US Nationwide in the states of WI, OH, NY, OH, IN, IA, TX, and countries of Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, UK
Recall number
Z-2561-2019
Recall status
Open, Classified
CDRH record ID
176161
FDA record created
2019-09-25
Recall event number
83692