Open original FDA record ↗Recall record
- FDA identifier
- Z-2561-2021
- Company in FDA record
- WOM World of Medicine AG
- FDA record creation date
- Sep 24, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When the display of inflow volume to the uterus reaches 32450 ml, the fluid management system control unit display will freeze because the internal software calculation threshold is reached. The outflow measurement will continue, and so the deficit accumulated up to this point will start counting backwards to 0 ml. There is a risk of distention fluid reaching the soft tissue circulatory system.
Source facts
- Action
- On 06/18/21 a recall notice was emailed to their one distributor customer. This customer was informed that a new software version was created that would alert the user when an inflow volume of 28.000 ml is reached. This is designed to prepare the user for manual determination of the deficit. The recalling firm recommends updating the units with the new software version during the next preventive maintenance at the recalling firm's U.S. location with the following contacts to coordinate shipment: Katrin.Glushko@novanta.com, 1-407-438-8810; Robert.Almeida@novanta.com, 1-407-472-1509
- Initiated
- 2021-06-18
- Root cause
- Software design
- Distribution
- US Nationwide distribution in the state of CT.
- Recall number
- Z-2561-2021
- Recall status
- Open, Classified
- CDRH record ID
- 188614
- FDA record created
- 2021-09-24
- Recall event number
- 88426