Open original FDA record ↗Recall record
- FDA identifier
- Z-2562-2011
- Company in FDA record
- Computerized Medical Systems Inc
- FDA classification date
- Jun 14, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Focal: If a plan involving Dynamic MLC Delivery is sent from XiO to Focal and called up there in Plan Review, the angle of the gantry is set to zero for all beams. If this plan is then sent from Focal to the R&V System, the gantry angles all remain at zero which could lead to patient mistreatment.
Source facts
- Action
- Elekta CMS Software sent a USER NOTICE dated November 5, 2010, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. A return postcard was included for the customer to confirm receipt of the User Notice. The firm intends to correct the software so it does not modify anything about the plan sent from XiO. Customers were instructed to contact their local customer support office with any questions.
- Initiated
- 2010-11-09
- Root cause
- Software design
- Terminated
- 2014-04-02
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Algeria, Australia, Belarus, Belgium, Brazil, Canada, Chile, China, Egypt, Estonia, Finland, France, Germany, Hong Kong, India, Italy, Latvia, Libya, Lithuania, Mexico, Morocco, Netherlands, New Zealand, Philippines, Poland, Romania, Singapore, Slovenia, South Africa, Spain, Suriname, Switzerland, Taiwan, Turkey, Turkmenistan, and the United Kingdom.
- Recall number
- Z-2562-2011
- Recall status
- Terminated
- CDRH record ID
- 97560
- FDA classified
- 2011-06-14
- Recall event number
- 57845